Autoren: G. De-Deus, V. Talarico Leal Vieira, E. J. Noqueira da Silva, H. Lopes, C. N. Elias, E. J. Moreira
Jahr: 2013
Journal: JOE
Rubriken: Cyclic fatigue
Schlüsselwörter:
Introduction The aim of the present study was to evaluate the bending resistance and the dynamic and static cyclic fatigue life of Reciproc R40 and WaveOne large instruments.
Methods A sample of 68 nickel-titanium instruments (25 mm in length) for use under reciprocation movement (Reciproc and WaveOne) from 3 different lots was tested. Reciproc R40 and WaveOne Large files, both of which had a nominal size of 0.40 mm at D0, were selected. The bending resistance was performed in 10 instruments of each system by using a universal testing machine. Dynamic and static models for cyclic fatigue testing were performed by using a custom-made device. For these tests, an artificial canal measuring 1.4 mm in diameter and 19 mm total length was fabricated from a stainless steel tube. Scanning electron microscopy analysis was performed to determine the mode of fracture. Statistical analysis was performed by using parametric methods, 1-way analysis of variance. Post hoc pair-wise comparisons were performed by using Tukey test for multiple comparisons.
Results WaveOne instruments presented significantly higher bending resistance than Reciproc (P < .05). Moreover, Reciproc revealed a significantly longer cyclic fatigue life (P < .05) in both static and dynamic tests (P < .05).
Conclusions Reciproc R40 instruments resisted dynamic and static cyclic fatigue significantly more than WaveOne Large instruments. Furthermore, WaveOne instruments presented significantly less flexibility than Reciproc.
Link zur Studie: 10.1016/j.joen.2013.10.013